The Beer–Lambert law, A = ε·c·l, says absorbance grows in proportion to how much absorbing material the light crosses: the molar extinction coefficient ε, the concentration c, and the path length l. This calculator solves that one equation for whichever quantity you are missing — absorbance, concentration, ε or path length — and it folds in the two bench terms most general converters ignore: the dilution you made before reading, and (with a molecular weight) the mass concentration in µg/mL beside the molar one. Enter transmittance instead of absorbance and it converts A = −log₁₀(T) for you, and it flags readings outside the band your spectrophotometer can actually trust.

One absorbance reading plus one extinction coefficient is enough to know how much of a compound is in a tube — that is the whole idea behind UV/vis quantitation, and it is the equation the biochemistry calculators are built on. The arithmetic is two keystrokes, yet the errors are remarkably consistent: people mix centimetres with millimetres, forget the dilution in front of the machine, take ε from a page that used a different solvent or path length, and trust an absorbance of 3 the detector never measured linearly. This page keeps every term of A = ε·c·l visible so nothing is silently assumed.

Absorbance itself is just a logarithm of how much light survived: A = −log₁₀(T), where T is the fraction of light transmitted. A reading of 1.0 means 10% of the light got through; 2.0 means 1%. The law then says that loss is proportional to concentration and to the distance the light travels through the sample, with ε as the proportionality constant — how strongly one mole per litre absorbs over one centimetre at your chosen wavelength. ε is where the chemistry lives: for NADH it is 6,220 L·mol⁻¹·cm⁻¹ near 340 nm, and that single number is what turns a 340 nm reading into a reaction rate.

Nucleic-acid and protein work usually hides this law inside conventions — the familiar 50 µg/mL-per-A260 rule is just ε compressed for a heterogeneous polymer, and it is what the DNA concentration calculator and the protein A280 calculator apply. This page is the general tool: you bring your own ε (from a reference or from a calibration curve you generate here by solving for ε) for any absorbing species at any wavelength, and the same dilution-and-path-length discipline applies.

Because most bench samples are diluted before they are readable, the concentration result is reported twice — what was in the cuvette, and what was in your tube before the dilution (cuvette × dilution factor). Supply a molecular weight and the answer also appears as a mass concentration in µg/mL, so an absorbance can become the number your protocol actually asks for. And whenever your input absorbance falls outside the roughly 0.01–2 band where instruments stay linear, the result comes with a warning rather than false confidence.

Beer-Lambert Law Calculator

Measured absorbance at the analyte’s wavelength maximum. Leave blank when solving for A.

Use instead of absorbance. Enter a fraction (0–1) or a percent, matching the selector beside it.

From a reference or a calibration — e.g. NADH is 6,220 L·mol⁻¹·cm⁻¹ near 340 nm.

In the unit chosen beside it. Leave blank when solving for c.

1 cm for a standard cuvette; short-path plates and microvolume readings differ.

Fold the sample was diluted before reading (concentration mode); multiplies the result.

Supply to also report the concentration in µg/mL and g/L.

Concentration100 µM
Absorbance0.622
Extinction coeff. (ε)6220 L·mol⁻¹·cm⁻¹
Cuvette concentration100 µM
Transmittance23.88 %T
Mass concentration—
Path length1 cm

Absorbance ↔ transmittance ↔ linear range

Absorbance is a logarithm of transmitted light, so each +1.0 absorbance is one factor of ten less light. Most bench spectrophotometers stay linear only over part of that scale — the middle of the band below is where a reading can be trusted.

Absorbance (A)Transmittance (T)%TReading quality
0.01.000100%At the blank — no analyte signal
0.10.79479.4%Low; watch the noise floor near 0.01
0.4340.36836.8%Most precise (≈1% photometric error)
1.00.10010%Upper end of the reliable range
2.00.0101%Stray light dominates — dilute and re-measure

The 0.01–2 band the calculator warns about is deliberately conservative; the classic “most precise” point near A = 0.434 is where the balance between noise and stray light is best. Exact limits depend on your instrument — treat the warning as a prompt to re-measure, not a verdict.

How to use the beer lambert law calculator

Choose what to solve for: absorbance, concentration, the extinction coefficient ε, or path length. The calculator shows only the inputs that equation needs.

Enter the measured absorbance — or enter transmittance instead (as a fraction or a percent, using the selector) and it converts with A = −log₁₀(T).

Enter ε in L·mol⁻¹·cm⁻¹ from a reference under similar solvent, pH and wavelength — for example 6,220 for NADH near 340 nm.

Enter concentration in whichever unit suits (M, mM, µM or nM) and path length in cm or mm; a standard cuvette is 1 cm.

In concentration mode, set the dilution factor if you diluted before reading — 10 for 1 µL sample into 9 µL buffer. The primary result is your original tube.

Optionally add a molecular weight to also see the concentration in µg/mL and g/L.

Press Calculate. Warnings appear automatically for readings outside the linear band; a note explains single-point ε when you solve for ε.

Ready to pipette? Carry the result into the dilution calculator or the molarity calculator, or follow the Beer–Lambert guide for the full workflow.

Beer-Lambert Law formula

Beer–Lambert law: A = ε · c · l (ε in L·mol⁻¹·cm⁻¹, c in mol/L, l in cm)
Solve for c: c = A / (ε · l) then × dilution factor for the original sample
Solve for ε: ε = A / (c · l) (a single-point calibration)
Solve for A: A = ε · c · l
Transmittance: A = −log₁₀(T) = 2 − log₁₀(%T)
Mass output: g/L = mol/L × MW; µg/mL = g/L × 1000
Absorbance is dimensionless; ε carries the wavelength in its identity — an ε at 260 nm is not an ε at 340 nm. All unit conversions (M↔µM, cm↔mm) happen inside the equation so the multiplication never silently mixes scales.

Variables and units

SymbolMeaningUnit
AAbsorbance at the chosen wavelength—
εMolar extinction coefficient (molar absorptivity)L·mol⁻¹·cm⁻¹
cConcentration of the absorbing speciesmol/L (or mM/µM/nM)
lPath length the light travels through the samplecm (or mm)
TTransmittance, fraction of light reaching the detector— (or %)
DFDilution factor applied before measuring×
MWMolecular weight (optional, enables mass output)g/mol

How beer lambert law is calculated

  1. Convert anything non-canonical into the equation’s units: µM→M, mm→cm, %T→A.
  2. Pick the missing quantity and apply the matching rearrangement of A = ε·c·l.
  3. For concentration, multiply the cuvette result by the dilution factor to get the original sample.
  4. For a mass figure, multiply mol/L by the molecular weight (g/L) and by 1,000 (µg/mL).
  5. Check the reading against the linear band (~0.01–2); outside it, dilute and re-measure rather than extrapolate.

Worked laboratory examples

Example 1 — NADH: absorbance to concentration

Given: A340 = 0.622 · ε = 6,220 L·mol⁻¹·cm⁻¹ · path length 1 cm

  1. c = A / (ε · l) = 0.622 / (6,220 × 1) = 1.00 × 10⁻⁴ mol/L.
  2. That is 100 µM — the reading goes straight to a concentration.
  3. With the molecular weight of NADH (664.4 g/mol) the same sample is 0.0664 g/L, i.e. 66.4 µg/mL.

Result: 100 µM (66.4 µg/mL)

Example 2 — The dilution you made before reading

Given: A340 = 0.622 · ε = 6,220 L·mol⁻¹·cm⁻¹ · 1 cm · 10 µL sample into 90 µL buffer (DF = 10)

  1. The cuvette contains 0.622 / 6,220 = 1.00 × 10⁻⁴ mol/L = 100 µM.
  2. Your tube is ten times stronger: 100 µM × 10 = 1.00 mM.
  3. Skipping the dilution factor is the most common way to report a sample ten-fold too weak.

Result: 1.00 mM original (100 µM in the cuvette)

Example 3 — Transmittance instead of absorbance

Given: T = 0.10 (10% of the light transmitted) · ε = 6,220 L·mol⁻¹·cm⁻¹ · 1 cm

  1. A = −log₁₀(0.10) = 1.000.
  2. c = 1.000 / 6,220 = 1.608 × 10⁻⁴ mol/L = 160.8 µM.
  3. At exactly one log of attenuation the reading is still inside the linear band; at T = 1% it would not be.

Result: 160.8 µM

Example 4 — Solving for ε (single-point calibration)

Given: A = 0.500 from a 20 µM standard · 1 cm path

  1. ε = A / (c · l) = 0.500 / (20 × 10⁻⁶ mol/L × 1 cm) = 25,000 L·mol⁻¹·cm⁻¹.
  2. Solve for path length just as easily: l = A / (ε · c) — handy when a short-path plate well is the unknown.
  3. One point carries that point’s error: a slope across several standards is sturdier, which the guide walks through.

Result: ε = 25,000 L·mol⁻¹·cm⁻¹

Want the reasoning behind the numbers? Read How to Use the Beer–Lambert Law.

How to interpret the result

The primary card answers exactly one question — the quantity you picked — and the cards beneath it keep the whole equation consistent: absorbance, ε, molar concentration (in the cuvette and in your original tube), transmittance, and mass concentration when you supplied a molecular weight. If two cards disagree with your mental math, the mistake is usually a unit that was never converted: µM entered as M is a 10⁶-fold error, and 10 mm entered as 10 cm is a 10-fold one.

The absorbance warning matters more than it looks. Above roughly 2, stray light reaching the detector makes readings compress — measured absorbance rises more slowly than concentration — so a 3.0 reading is not three times a 1.0 reading. Between 1 and 2 the deviation is instrument-dependent, and the calculator notes it. Below about 0.01 you are comparing a signal to the blank’s drift. The fix in both directions is sample preparation, not arithmetic: dilute toward the 0.1–1 band, or concentrate the sample.

When you solve for ε, the result is a single-point calibration coefficient. It is honest for checking a reference value and fragile for everything else: one contaminated standard silently rescales every future concentration that uses it. Three or more standards in the linear band and a slope give the same ε with the scatter visible.

Scattering is the silent bias no equation can see. Turbid suspensions, bubbles and particulates deflect light out of the beam, and the detector reports the loss as absorbance — the reading inflates at all wavelengths even though nothing “absorbs”. For turbid samples the convention is to blank against the same matrix and keep the wavelength away from the analyte’s peak when checking the background.

Assumptions

  • A single absorbing species dominates at the measurement wavelength.
  • The solution is dilute enough that molecules do not interact (deviations start around 0.01–0.02 M for many dyes).
  • Monochromatic light at the wavelength ε refers to; broadband instrument settings shift effective ε.
  • The blank zeroed the solvent and cuvette contributions; the path length entered is the one the light actually travelled.
  • Dilution factor is folded in only in concentration mode, where it has a meaning.

Limitations

  • Near absorbances above ~2 results carry systematic error no rearrangement can fix; treat them as estimates pending a re-measure.
  • ε depends on solvent, pH, temperature and ionic strength — take it from a source matched to your conditions or measure it.
  • The tool is molar (ε-based); the nucleic-acid and protein conventions with their fixed per-A260/A280 factors live in their own calculators.
  • No background correction: the calculator assumes the instrument’s blank already removed solvent and cuvette absorption.

Common mistakes

  • Forgetting the dilution factor, so the reported concentration is of the cuvette, not the tube.
  • Using ε from one wavelength or solvent with a reading taken at another.
  • Entering a 10 mm micro-volume path as 10 cm — a 10-fold error hiding in a unit.
  • Trusting absorbances above ~2 instead of diluting into the linear band.
  • Reading transmittance where absorbance is expected (or vice versa) — always check which scale your instrument outputs.
  • Using A260 conventions for a purified small molecule, or a molecular ε for a heterogeneous polymer.

FAQ

How do you calculate concentration from absorbance?

Divide the absorbance by the product of extinction coefficient and path length: c = A / (ε·l), in mol/L when ε is in L·mol⁻¹·cm⁻¹ and l in cm. If you diluted the sample before measuring, multiply by the dilution factor to get the original concentration. This calculator does the conversions and the warning checks in one step.

What are the units of the molar extinction coefficient?

Litres per mole per centimetre (L·mol⁻¹·cm⁻¹, often written M⁻¹cm⁻¹) — because absorbance is dimensionless, ε must cancel concentration × path length. A mass-based cousin, A(1%, 1 cm), uses different units; for protein A280 conventions use the dedicated protein concentration calculator.

How is transmittance related to absorbance?

A = −log₁₀(T), where T is the fraction of light that passes through, so 100% T is A = 0, 10% T is A = 1 and 1% T is A = 2. Enter either scale — fraction or percent — in place of absorbance and the calculator converts it.

Why is the path length 1 cm?

Because the classic cuvette is 1 cm and published ε values are usually normalized to it. The law needs the actual distance the light crossed the liquid: micro-volume instruments use fractions of a millimetre, and plate readers use whatever the well volume gives you. Enter the true path (cm or mm); the calculator uses it, and you can solve for it when it is the unknown.

When does the Beer-Lambert law stop being linear?

Mainly in four situations: at high absorbance where stray light dominates (roughly above 1–2), at high concentration where molecules interact, when the light is not specific enough in wavelength, and when the sample scatters light rather than absorbing it. The warning band on this tool covers the first; the assumptions section states the rest.

Can I find an extinction coefficient with this calculator?

Yes — choose ε as the solve target and enter a reading with its known concentration and path length. That is a single-point calibration. For real work measure several standards and take the slope of absorbance against concentration; the guide shows how, and one point alone inherits its own measurement error.

Is this the same as the A260 DNA calculator?

Same law, different bookkeeping. The DNA concentration calculator uses the field’s conventional factors (the “50 µg/mL per A260” style rules) because nucleic acids are heterogeneous polymers. When you have a defined compound with a real ε at any wavelength — cofactors, dyes, drug candidates — this general solver is the right tool.

Scientific Sources

External links open in a new tab and carry rel="nofollow". We cite primary references (NIST, NIH/NCBI, peer-reviewed papers) and established manufacturer technical documentation (NEB, Thermo Fisher, Promega, IDT). Calculated results are estimates; your protocol and manufacturer guidance take precedence.